American West Analytical Laboratories

    Home Up Contact AWAL Contents Search

Soil Organics
      

    NEWS  References  Capabilities  About AWAL

Solid / Soil Matrix2
Analyte Method Sample Quantity1,2 Container Type Preservation Hold Time
Organics      
Volatiles by GC/MS SW846 8260C * * * 14 days
Semivolatiles by GC/MS SW846 8270D 30 g amber glass/Teflon cool 4oC 14 days ext./ 40 days analysis
Organochlorine Pesticides by GC/ECD SW846 8081B 30 g amber glass/Teflon cool 4oC 14 days ext./ 40 days analysis
Herbicides by GC/ECD SW846 8151A 30 g amber glass/Teflon cool 4oC 14 days ext./ 40 days analysis
Diesel Range Organics by GC/FID SW846 8015D Modified 30 g amber glass/Teflon cool 4oC 14 days ext./ 40 days analysis
PCBs by GC/ECD SW846 8082A 30 g amber glass/Teflon cool 4oC 14 days ext./ 40 days analysis
TPPH by GC/MS SW846 8260C * amber glass/Teflon cool 4oC 14 days
TPH (GRO) (C5-C10) Utah by GC/MS SW846 8260C * amber glass/Teflon cool 4oC 14 days
TPH (DRO) (C11-C15) by GC/MS SW846 8260C * amber glass/Teflon cool 4oC 14 days
TPH (DRO) Utah by GC/PID&FID SW846 8015D Mod. * amber glass/Teflon cool 4oC 14 days
VOA TPH Fractionation by GC/MS (Utah) SW846 8260C * amber glass/Teflon cool 4oC 14 days
SVOA TPH Fractionation by GC/MS (Utah) SW846 8270D 30 g amber glass/Teflon cool 4oC 14 days ext./ 40 days analysis
TPH (DRO) Non-Utah SW846 8015D modified 30 g amber glass/Teflon cool 4oC various
           
           
* If sampling by method 5030 (not preferred) collect the sample in a 4 oz. glass jar with no headspace then cool to 4oC.          
* If sampling by method 5035 (preferred) collect a 5 gram sample for low level analysis in each of 3 - 40ml glass VOA vials with Teflon lined septa then freeze the samples ASAP. For high level analysis collect a 5 gram sample in each of two preweighed 40 ml VOA vials that have 10 ml of methanol already in the vial then cool to 4oC.
           
1 - This does not include the volume necessary to perform MS/MSD analyses. Include additional sample (normally 3 times the quoted volume or mass) for each sample requiring an MS/MSD. This extra sample volume should always be included for at least one sample in sample sets of 5 or greater. Contact the laboratory for more information.
2 - For reporting results in dry weight a total solid analysis must be included. Be sure to collect a sample for total solids.

 

 

Last modified: December 15, 2007